A manometer is an instrument for measuring pressure or the difference between two pressures. In its traditional and narrower sense, it uses a liquid column whose height changes until the liquid’s weight balances the applied pressure difference. The term is also used more broadly for mechanical and electronic pressure-measuring instruments, although these are often called pressure gauges. Manometers are used in laboratories, ventilation systems, industrial equipment, and instrument calibration.
Measurement principle
A liquid-column manometer operates according to hydrostatics, the study of fluids at rest. In a stationary liquid of uniform density, the pressure difference between two levels is
where is the liquid density, is the local acceleration due to gravity, and is the vertical separation between the levels. Pressure increases with depth.
In a simple U-shaped tube partly filled with liquid, pressure applied to either arm moves the liquid until equilibrium is reached. If the pressures above the two liquid surfaces are and , and the second surface stands a vertical distance above the first, then, neglecting the weight of the gas above the liquid,
The measured quantity is the difference between the surface elevations, not the displacement of just one surface from its initial position. In a uniform-bore U-tube, a rise in one arm is accompanied by an equal fall in the other.
When the pressure-transmitting fluid has appreciable density, its weight must also be included. For a differential U-tube containing a denser manometric liquid, with pressure connections at the same elevation and the same lighter fluid above both interfaces,
where is the manometric-liquid density and is the density of the fluid being measured. For gases near ordinary atmospheric conditions, the density correction is often small; for liquids, it can be substantial. Connections at different elevations require additional hydrostatic corrections.
Pressure references
A manometer’s reading depends on the reference pressure used.
- Gauge pressure is measured relative to the surrounding atmosphere. An open-ended liquid manometer has one arm exposed to atmospheric pressure.
- Differential pressure is the difference between two connected pressure sources, such as locations upstream and downstream of a filter.
- Absolute pressure is measured relative to a vacuum. A sealed reference space must have a sufficiently low and known residual pressure for accurate absolute measurement.
These quantities are related by
Negative gauge pressure indicates pressure below the surrounding atmosphere, not negative absolute pressure.
A barometer is a related instrument specifically intended to measure atmospheric pressure. A mercury barometer can be understood as a liquid-column instrument in which atmospheric pressure supports a column against a low-pressure space above it.
Main types
U-tube manometer
The U-tube is the simplest common design. A transparent tube holds a liquid such as water, oil, or mercury, and a scale indicates the vertical difference between its two surfaces.
Its advantages include simple construction, a readily understood operating principle, and no need for electrical power. Its disadvantages include limited portability, susceptibility to spilling, and the length of column required for large pressure differences. A pressure difference of one standard atmosphere supports approximately 10.3 metres of water, compared with about 0.76 metres of mercury.
Well-type manometer
A well-type, or cistern, manometer connects a narrow measuring tube to a reservoir with a much larger cross-sectional area. Most of the visible level change occurs in the narrow tube, allowing the instrument to be read against a single scale.
The reservoir level nevertheless changes slightly. Accurate readings therefore require a correction or a scale designed to account for the relative cross-sectional areas. Treating the reservoir level as perfectly fixed is an approximation.
Inclined manometer
An inclined manometer places the measuring tube at a shallow angle to the horizontal. A small vertical change then produces a larger displacement along the tube, making small pressure differences easier to read.
For a tube inclined at angle ,
where is the displacement along the tube corresponding to the vertical height . In a practical well-type instrument, reservoir movement must also be included. Inclination improves readability; it does not change the hydrostatic pressure associated with a given vertical height.
Inclined instruments are particularly useful for low differential pressures in ducts and ventilation systems.
Differential and inverted manometers
A differential manometer connects both sides to pressure sources rather than leaving one side open to the atmosphere. It can measure a small difference even when both sources are at a substantial common pressure, provided the instrument and its connections can withstand that pressure.
An inverted U-tube traps a lighter fluid above a denser process liquid. This arrangement is useful for measuring small differences between liquid pressures. Its calculation depends on the densities and elevations of the relevant fluid interfaces.
Mechanical and electronic instruments
In broader usage, manometer includes instruments that do not use a visible liquid column.
A Bourdon-tube gauge uses a curved, flattened tube that tends to straighten as internal pressure increases. A linkage converts its movement into pointer rotation. Diaphragm and bellows instruments similarly use elastic deformation.
Electronic instruments employ a pressure-sensitive sensor, commonly based on piezoresistive or capacitive effects. Many use microelectromechanical systems to form a small diaphragm and its sensing elements. Their outputs may be displayed locally, recorded, or transmitted to a process-control system.
Unlike a simple liquid column, these instruments infer pressure from a calibrated relationship between pressure and deformation or electrical response.
Manometric liquids and units
The choice of liquid affects measuring range, sensitivity, and compatibility.
Water is inexpensive and provides a relatively large displacement for a small pressure difference. Oils offer a range of densities and wetting properties. Mercury provides a compact column because of its high density and has low vapour pressure at ordinary temperatures, but its toxicity and environmental hazards limit its use.
Other important properties include viscosity, evaporation rate, chemical stability, and interaction with the tube and process fluid. Viscosity primarily affects settling time rather than the final hydrostatic balance.
The International System of Units expresses pressure in pascals:
Liquid-column units, including millimetres of mercury and inches of water, remain in use. Their conversion requires attention to the specified convention. An actual column’s pressure depends on liquid density, temperature, and local gravity, whereas standardized pressure units based on liquid columns use defined reference conditions.
History
Liquid-column pressure measurement developed from seventeenth-century investigations of atmospheric pressure and vacuum. Evangelista Torricelli’s mercury experiment in 1643 demonstrated that atmospheric pressure could support a column of liquid, establishing the operating principle of the mercury barometer.
As experimental science and industrial machinery expanded, liquid manometers became important for comparing gas pressures and investigating fluid flow. The growth of steam engines and boilers also created demand for compact, robust instruments capable of measuring higher pressures without very tall liquid columns.
Eugène Bourdon patented his elastic-tube pressure gauge in 1849. Mechanical gauges subsequently became widespread in industrial service. Electrical transducers and semiconductor fabrication later enabled remote indication, automatic recording, and digital pressure instruments. Liquid-column manometers have remained useful where simplicity and a direct hydrostatic measurement are valuable.
Applications
Manometers serve several distinct measurement tasks:
- Ventilation and air handling: measuring duct pressure, fan pressure differences, and pressure loss across filters in air-conditioning systems.
- Flow measurement: measuring the pressure difference produced by an orifice plate, Venturi tube, or Pitot-static arrangement. Flow is then inferred using the appropriate fluid-flow relationship and instrument calibration.
- Laboratory work: comparing gas pressures and studying hydrostatic and fluid-flow behaviour.
- Industrial monitoring: observing pressures and pressure losses in vessels, pipelines, and equipment used in chemical engineering.
- Calibration: comparing other pressure instruments with a liquid-column reference under controlled conditions.
Pressure-drop measurement does not by itself determine flow rate. The relationship also depends on geometry, fluid density, and the applicable flow regime.
Accuracy and limitations
A liquid manometer has a simple physical basis, but its accuracy depends on more than reading a scale.
Density and temperature. Liquid density changes with temperature. Temperature also affects the dimensions of the tube and scale, making thermal conditions relevant in precise work.
Meniscus and capillary effects. Surface tension curves the liquid surface and can cause capillary rise or depression. Unequal tube diameters or different wetting conditions may produce systematic offsets.
Alignment and reading. Vertical instruments must be levelled appropriately, while inclined instruments require a known angle. Parallax, meniscus interpretation, and limited scale resolution contribute to measurement uncertainty.
Fluid condition and connections. Bubbles, leaks, contamination, evaporation, and trapped liquid in pressure lines can alter readings. Pressure-line elevations matter when the connecting fluid’s weight is significant.
Dynamic response. Liquid inertia and viscosity delay the response to changing pressure. Rapid fluctuations may produce oscillation rather than a stable indication; damping improves readability but slows response.
Range and containment. Large pressure differences require tall columns or dense liquids. Excessive differential pressure can expel the manometric liquid, while high common pressure requires suitably rated tubes, fittings, and seals.
Mechanical and electronic instruments avoid several liquid-column limitations but introduce others, including hysteresis, temperature sensitivity, zero drift, and long-term changes in calibration. A digital display with many digits does not, by itself, establish high measurement accuracy.