Burette

A burette is a calibrated laboratory instrument that dispenses measured, variable volumes of liquid, chiefly for titration in quantitative chemical analysis.

15 keywords1 linked from1 not yet writtenWritten by AI
TitrationAnalytical Chemi…MeniscusSurface TensionCalibrationpHStoichiometryMolar Concentrat…Burette

A burette is a laboratory instrument used to deliver accurately measured, variable volumes of liquid. Its most familiar form is a long, graduated glass tube with a stopcock and a narrow delivery tip at the bottom. The volume dispensed is determined from the difference between two scale readings. Burettes are central to titration, a method of analytical chemistry in which a measured quantity of reagent is added to a sample to determine the amount or concentration of a substance.

Construction and operating principle

A conventional liquid burette consists of a cylindrical tube, a graduated scale, and a flow-control valve. The tube is usually made from chemically resistant glass. The stopcock may be a ground-glass plug or incorporate a chemically resistant polymer such as polytetrafluoroethylene (PTFE). A tapered outlet allows liquid to be added as a stream or as individual drops.

The scale normally increases downward: zero is near the top, and progressively larger readings indicate that more liquid has left the tube. For an initial reading RiR_i and a final reading RfR_f, the indicated volume delivered is

V=Rf−Ri.V = R_f-R_i.

The instrument does not need to begin at exactly zero. What matters is that both readings are taken correctly within the calibrated range and that the delivery tip remains filled with liquid.

Common laboratory capacities include 10, 25, and 50 mL. A conventional 50 mL burette often has graduations at 0.1 mL intervals. Readings may be estimated between graduations, but the number of recorded digits must not be confused with the instrument’s actual accuracy.

Unlike a pipette designed to transfer a fixed volume, a burette can deliver a continuously adjustable quantity while allowing the operator to measure the amount dispensed.

Reading the liquid level

The surface of a liquid in a narrow tube usually forms a curved meniscus because of the interactions between the liquid and the tube wall, together with surface tension. For a clear aqueous solution in clean glass, the usual convention is to read the lowest point of the concave meniscus.

The eye must be level with the meniscus. Looking from above or below produces parallax error, in which the apparent position of the liquid surface shifts relative to the scale. A contrasting background can make the meniscus easier to identify.

For strongly colored liquids whose lower meniscus cannot be seen clearly, a different reading convention may be specified. The same convention must be used for both readings and must be consistent with the applicable analytical method.

Some burettes have an optical stripe behind the tube that makes the meniscus easier to locate. This improves readability but does not replace correct alignment or calibration.

Use in titration

In titration, the burette normally contains the titrant, a reagent solution of known concentration. The sample, or analyte, is held in a separate vessel and mixed as the titrant is added.

The endpoint may be detected by a color-changing indicator or by an instrument, such as a pH meter. The observed endpoint is distinct from the equivalence point: the equivalence point is the condition at which the reactants have combined in the proportions specified by the reaction’s stoichiometry. A difference between the two can introduce analytical error.

For a reaction

aA+bB→products,aA+bB\rightarrow \text{products},

where AA is the analyte and BB the titrant, the equivalence condition is

nAa=nBb.\frac{n_A}{a}=\frac{n_B}{b}.

If the titrant has molar concentration cBc_B and the corrected volume required is VBV_B, then

nA=abcBVB.n_A=\frac{a}{b}c_BV_B.

Here, concentration and volume must be expressed in compatible units. For a simple 1:1 reaction with no additional corrections,

cAVA=cBVB.c_AV_A=c_BV_B.

Burettes are used in acid–base titrations, oxidation–reduction titrations, precipitation titrations, and complexometric titrations. Applications include determining acidity, measuring water hardness, and assaying chemical products.

Conditions for reliable delivery

Several conditions are important for valid measurements:

  • Clean internal surfaces: Liquid should drain uniformly rather than leave isolated droplets on greasy or contaminated glass.
  • Conditioning with titrant: Residual rinse water can dilute the reagent. The burette is therefore commonly rinsed with small portions of the solution it will contain.
  • A filled delivery tip: An air bubble can cause part of the indicated volume change to fill the tip rather than reach the sample.
  • A leak-free stopcock: Leakage can change the reading without the intended addition to the receiving vessel.
  • Controlled addition and mixing: Near the endpoint, small additions help distinguish the endpoint without excessive overshoot.
  • Consistent drainage and reading: Liquid remaining on the walls takes time to drain. Applicable waiting times and delivery procedures must be observed.

A hanging drop on the delivery tip can also matter. Standard methods commonly specify how liquid at the tip is transferred to the receiving vessel without washing additional liquid out of the burette.

Types

Graduated glass burettes

These are the conventional manually operated instruments. Different capacities allow the delivered volume to be matched to the scale of the analysis. Smaller-capacity burettes can provide finer volume discrimination, although performance depends on their design and calibration.

Microburettes

Microburettes dispense small quantities, often for analyses involving limited samples or low reagent consumption. Their designs may use narrow graduated tubes or precision piston mechanisms. Small delivered volumes make evaporation, bubbles, and liquid retained at the outlet especially significant.

Piston and digital burettes

A piston burette dispenses liquid by moving a piston through a cylinder. Its delivery volume is determined from piston displacement rather than from the difference between liquid levels in an open graduated tube.

Digital burettes display the dispensed volume electronically or through a mechanical counter. Many are mounted directly on reagent bottles. Their performance depends on the cylinder geometry, drive mechanism, seals, valve operation, and calibration; a high-resolution display alone does not establish accuracy.

Automatic titration systems

Automatic titrators combine a motor-driven burette with a detector and control system. They can adjust the addition rate, record a titration curve, and calculate an endpoint. Automation can improve consistency and throughput, but it does not remove errors from unsuitable reaction conditions, incorrect reagent concentration, bubbles, or an inappropriate endpoint-detection method.

Calibration and measurement uncertainty

Conventional burettes are generally calibrated to deliver, rather than simply to contain, a stated volume. Markings such as “TD” or “Ex” indicate this delivery basis. Calibration accounts for liquid retained on the walls under specified operating conditions.

Calibration commonly uses a gravimetric method: the dispensed water is weighed, and its volume is calculated from its density at the measured temperature. More precise work also accounts for effects such as air buoyancy. Checks at several positions can reveal variations in the scale rather than merely establish the total capacity.

Volumetric glassware is often classified by permissible error, with Class A generally having tighter tolerances than Class B under the relevant standard. Exact tolerances depend on the capacity and specification. Some standards also distinguish instruments with defined delivery and waiting-time requirements.

The measurement uncertainty of a delivered volume includes more than scale readability. Contributions may arise from calibration, reading repeatability, drainage, leakage, bubbles, temperature, and operator technique. Since a conventional burette measurement uses two readings, both contribute to the uncertainty of their difference. Calibration errors at different scale positions may be correlated, so they cannot always be treated as independent contributions.

Development and limitations

Burettes became important during the nineteenth-century development of volumetric chemical analysis. Early flow-control arrangements included pinched flexible tubing; glass stopcocks subsequently became widely used. Later developments introduced chemically resistant polymer components, piston-driven delivery, digital readouts, and automated control.

A burette measures delivered volume, not chemical composition directly. The validity of an analytical result also depends on reagent concentration, reaction completeness, selectivity, and endpoint detection. Volatile liquids, unusually viscous solutions, and liquids that wet the apparatus differently from the calibration liquid can require special procedures or different dispensing equipment. An instrument calibrated with water does not necessarily deliver every liquid with the same error under otherwise unchanged conditions.