Thermogravimetric analysis (TGA) is a technique in thermal analysis that measures a sample’s mass as a function of temperature or time while the sample follows a controlled temperature program in a specified atmosphere. Mass changes can reveal the loss of moisture or volatile substances, thermal decomposition, and reactions that incorporate or release gases. Widely used in analytical chemistry and materials characterization, TGA provides quantitative information about mass changes but generally requires complementary measurements to identify the substances responsible.
Measurement principle
A thermogravimetric instrument continuously weighs a sample while controlling its thermal environment. The principal output is a thermogravimetric, or TG, curve: a plot of mass, often expressed as a percentage of the initial mass, against temperature or time.
If the initial sample mass is and its measured mass at time is , the percentage mass remaining is
The percentage mass loss is correspondingly
Mass loss may result from evaporation, release of adsorbed gases, removal of chemically bound water, or decomposition into volatile products. Mass gain may result from oxidation or another reaction with the surrounding gas. Competing processes can occur simultaneously, so the observed curve represents their net effect.
TGA measures mass change rather than heat flow. A phase transition such as melting may produce no direct TG signal if no material enters or leaves the sample. Likewise, substantial chemical or structural changes can occur while the measured mass remains nearly constant.
Instrumentation and experimental conditions
A typical thermogravimetric analyzer contains:
- A sensitive balance, which records the mass of the sample and its holder.
- A furnace, which applies the temperature program.
- A sample pan or crucible, selected for chemical compatibility and the required temperature range.
- A temperature sensor, positioned to characterize the sample’s thermal environment.
- A gas-control system, which establishes the atmosphere and removes evolved products.
- Data-acquisition and control software, which coordinates measurements and calculates quantities such as mass-loss rates.
Samples commonly weigh a few milligrams to a few tens of milligrams, although appropriate masses depend on the instrument and material. Small, evenly distributed specimens generally reduce temperature gradients and restrictions on gas transport. Larger samples can improve the detectability of small composition fractions but may introduce thermal lag, self-heating, or incomplete removal of volatile products.
Crucibles may be made from alumina, platinum, or other materials. Crucible composition, dimensions, lid configuration, and contact with the sample can influence results. A crucible must not react significantly with the specimen or its decomposition products.
Atmosphere and pressure
The gas surrounding the sample is part of the experimental definition, not merely an instrumental detail. Nitrogen and argon are commonly used to provide nominally inert conditions. Air or oxygen permits oxidative reactions and combustion. Other gas compositions can be used to study reduction, carbonation, or reactions under process-relevant conditions.
A nominally inert gas is not chemically inert toward every material. Trace oxygen, moisture, and other impurities may affect sensitive samples.
Gas flow influences the removal of volatile products, their partial pressures near the sample, and heat transfer. Pressure can also change evaporation rates and reaction behavior. Consequently, measurements made under different flow rates, gas compositions, or pressures are not automatically comparable.
Temperature programs
Three common operating modes are:
- Dynamic or non-isothermal TGA: temperature changes at a specified rate, usually during heating.
- Isothermal TGA: the sample is held at a selected temperature while mass is recorded over time.
- Stepwise or controlled-rate TGA: the program uses temperature steps or adjusts heating in response to the observed mass-loss rate.
A single experiment may combine several modes or switch atmospheres. For example, heating a polymer composite first in nitrogen and then in air can help distinguish volatile decomposition products, combustible carbonaceous residue, and noncombustible inorganic residue.
Reading thermogravimetric curves
Mass-loss steps and plateaus
A TG curve may show one or several mass-loss steps separated by plateaus. Each step represents a temperature or time interval over which the sample undergoes a net mass change. A plateau indicates little measurable net change under the conditions of that interval; it does not establish the absence of all chemical reactions.
Possible interpretations include:
| Observed feature | Possible explanation |
|---|---|
| Low-temperature mass loss | Loss of moisture, residual solvent, or other volatile material |
| Distinct intermediate-temperature step | Dehydration, decomposition, or release of a particular component |
| Broad or overlapping loss | Several concurrent reactions or a distribution of decomposition processes |
| Mass gain in an oxidizing atmosphere | Uptake of oxygen by the sample |
| Residue after heating | Inorganic material, char, unreacted sample, or a mixture |
These assignments are hypotheses until supported by the sample’s chemistry or independent evidence. A low-temperature loss, for example, is not necessarily water.
Derivative thermogravimetry
Derivative thermogravimetry (DTG) plots the rate of mass change against temperature or time. Common forms are and . For a constant positive heating rate ,
Depending on the plotting convention, mass loss appears as a negative trough or as a positive peak after reversing the sign. DTG curves make changes in decomposition rate easier to recognize and may reveal shoulders that are inconspicuous on the original TG curve.
A DTG peak marks the maximum observed mass-loss rate within a process or group of overlapping processes. It is not necessarily the temperature of a single elementary reaction. Numerical differentiation also amplifies noise, and smoothing can change apparent peak shape and separation.
Characteristic temperatures
Results are often summarized using an onset temperature, the temperature at a specified percentage loss, or the temperature of maximum mass-loss rate. These quantities are not interchangeable.
An extrapolated onset is commonly determined by intersecting a baseline with a tangent to a steep region of the TG curve. A temperature at 5% mass loss instead depends on the initial mass reference and any earlier losses. Both depend on the experimental conditions and calculation procedure. A quoted “decomposition temperature” is therefore incomplete unless its operational definition is stated.
Quantitative composition analysis
TGA can determine composition when a mass change has a known and sufficiently selective chemical interpretation. The calculation relies on stoichiometry, complete conversion, and the absence of interfering processes.
For example, calcium carbonate can decompose according to
Using approximate molar masses, the theoretical mass loss from pure calcium carbonate is
If a mixture loses a fraction of its initial mass exclusively through this reaction, its initial calcium carbonate mass fraction is
The inference fails if other components release gases in the same interval, if conversion is incomplete, or if concurrent mass gain offsets the loss. The carbon dioxide partial pressure also affects carbonate decomposition conditions.
Similar approaches can quantify water of crystallization, volatile content, and selected filler or residue fractions. Final residue should not automatically be equated with the original inorganic content: inorganic constituents may themselves decompose, volatilize, or gain mass during heating.
Reaction kinetics
TGA is widely used to investigate chemical kinetics, particularly decomposition and gas–solid reactions. For a well-defined mass-loss process, the extent of conversion may be written as
where and are the initial and final masses assigned to that process.
A common kinetic representation is
with an Arrhenius temperature dependence,
Here, is a pre-exponential factor, is an apparent activation energy, is the gas constant, and describes an assumed conversion dependence.
Experiments at multiple heating rates or temperatures are substantially more informative than a single heating curve. Isoconversional methods compare rates or temperatures at equal conversion across experiments and estimate apparent activation energies without imposing one fixed conversion function over the whole process.
Kinetic parameters require cautious interpretation. Overlapping reactions, changes in mechanism, diffusion limitations, and temperature gradients can all influence them. A good fit to one curve does not establish a unique reaction mechanism. Extrapolating short high-temperature experiments to long-term service conditions can be unreliable if the governing processes change.
Coupled and complementary techniques
TGA records how much mass changes, but not directly which substances leave the sample. Coupled measurements help resolve this limitation.
- TGA–mass spectrometry: evolved gases are transferred to a mass spectrometer, which records ion signals associated with volatile products. Fragmentation and overlapping signals can complicate identification.
- TGA–infrared spectroscopy: evolved gases are examined through their infrared absorption bands. This is useful for many molecular gases but does not directly detect infrared-inactive species such as nitrogen.
- TGA with gas chromatography: evolved components can be separated before detection, improving identification in complex mixtures.
- Simultaneous TGA and differential scanning calorimetry: mass change and heat flow are measured on the same specimen under a shared temperature program. This distinguishes mass-changing events from thermal events without appreciable mass change.
Transfer lines in evolved-gas systems must limit condensation and secondary reactions. Gas transport also introduces a delay between mass change at the sample and detection downstream.
Residues can be examined by X-ray diffraction, microscopy, or other forms of spectroscopy to identify solid products.
Applications
Polymers and composites
TGA characterizes decomposition behavior, residual solvent, volatile additives, and residue formation in polymers. Controlled changes of atmosphere can help separate polymer decomposition, carbonaceous residue oxidation, and inorganic filler contributions. These measurements support formulation comparison and quality control, but a single mass-loss temperature is not a comprehensive measure of service lifetime.
Minerals and inorganic materials
Minerals, ceramics, and cementitious materials may undergo dehydration, dehydroxylation, carbonate decomposition, or oxidation during heating. Their mass changes can help quantify selected phases when assignments are supported by complementary analysis.
Fuels and biomass
TGA is used to investigate drying, volatile release, char formation, and oxidation in coal, biomass, and other fuels. Constituents such as cellulose and lignin have overlapping thermal behavior, so separating their contributions solely from curve shape is generally model-dependent.
Sorbents and reactive materials
Mass uptake and release can characterize adsorption, regeneration, oxidation, and reduction. Repeated atmosphere or temperature cycles are useful for assessing reversible capacity and changes during cycling.
Limitations, calibration, and reporting
TGA curves depend on both the material and the measurement procedure. Important sources of variation include:
- Heating rate: faster heating commonly shifts observed events to higher temperatures because less time is available for conversion at each temperature and thermal lag may increase.
- Heat and mass transfer: sample size, packing, and crucible geometry can control apparent behavior.
- Buoyancy and gas-flow effects: changing gas density and flow can produce apparent balance changes unrelated to sample chemistry.
- Baseline drift and contamination: instrument drift or deposits can distort small signals.
- Sample heterogeneity: a small specimen may not represent a bulk mixture.
- Overlapping processes: several reactions can produce a seemingly simple step.
- Limited chemical specificity: identical mass losses can arise from different substances or mechanisms.
Reliable work requires appropriate calibration of mass and temperature, baseline assessment, and consideration of measurement uncertainty. Blank runs can help characterize instrumental contributions, although they do not reproduce every sample-dependent effect.
A reproducible report specifies sample preparation and initial mass, crucible type, atmosphere and flow, pressure where relevant, temperature program, and the definitions used for onset, mass-loss intervals, and residue. Comparisons of thermal stability are meaningful only when these conditions and reporting conventions are sufficiently consistent.