Image-guided radiation therapy (IGRT) is the use of imaging immediately before or during radiation therapy to locate the treatment target, assess surrounding anatomy, and guide adjustments to treatment delivery. Most commonly used in external-beam radiation therapy for cancer, it helps account for differences between the anatomy represented in the treatment plan and the anatomy present at a treatment session. IGRT is a method of treatment guidance rather than a separate type of radiation.
Purpose and principles
Radiation treatment is planned using images that define the target and nearby normal tissues. However, a patient’s position and internal anatomy cannot be reproduced perfectly at every session. Tumors and organs may move with breathing, bladder filling, bowel activity, swallowing, or changes in body shape. Tumor shrinkage and weight loss can also alter anatomy over a course of treatment.
IGRT addresses two broad sources of variation:
- Interfraction variation: differences between treatment sessions, such as changes in positioning or organ filling.
- Intrafraction motion: movement during a session, including respiratory motion and gradual displacement after positioning.
Treatment images are compared with reference images through image registration. Depending on the system and clinical protocol, the comparison may lead to a treatment-table shift, correction of patient rotation, repositioning, interruption of treatment, or reassessment of the treatment plan.
Planning margins allow for uncertainties in target position and radiation delivery. Image guidance can reduce some of these uncertainties and may permit smaller margins in appropriate settings. It does not eliminate all uncertainty: image quality, target visibility, movement after imaging, and mechanical accuracy remain relevant.
Imaging methods
X-ray imaging and cone-beam CT
Two-dimensional X-ray images can show bones or implanted markers that serve as landmarks. Images obtained from different directions help determine positional differences. Some systems use megavoltage imaging produced with the treatment beam, while others use a separate kilovoltage X-ray source.
Cone-beam computed tomography acquires projections around the patient and reconstructs a three-dimensional image. It allows assessment of internal anatomy in the treatment position and is widely integrated into treatment machines. Compared with diagnostic computed tomography, image quality may be more affected by scatter, motion, and limited soft-tissue contrast. Other arrangements use a CT scanner in the treatment room or integrated with the treatment platform.
MRI guidance
Magnetic resonance imaging provides soft-tissue contrast without an additional ionizing imaging exposure. Systems combining MRI with radiation delivery can visualize selected targets and nearby organs before treatment and, in some configurations, during irradiation.
MRI guidance can support motion monitoring and treatment adaptation. Its limitations include geometric distortion, restrictions associated with implants or other MRI safety considerations, and technical effects of the magnetic field on radiation dose distribution.
Ultrasound and surface guidance
Ultrasound can localize selected soft-tissue structures without ionizing radiation. Its performance depends on image quality, operator technique, and the reproducibility of probe placement; probe pressure can also change anatomy.
Optical surface-guidance systems track the external body surface. They can support positioning, breath-hold monitoring, and detection of movement. However, the surface is a surrogate for internal anatomy and does not directly establish the position of every internal target.
Treatment workflow
A typical image-guided treatment session includes:
- Initial positioning: the patient is placed in the planned treatment position, often using immobilization devices and external alignment references.
- Image acquisition: images are obtained using the selected guidance system.
- Comparison and review: current images are aligned with reference images, with attention to the target and relevant nearby tissues.
- Correction and verification: positional corrections are applied and checked as required.
- Radiation delivery: treatment proceeds, with additional monitoring or imaging when the technique requires it.
Not every image-guidance system continuously tracks the tumor. A pretreatment image verifies anatomy at a particular time; subsequent movement may still occur.
Implanted markers, often called fiducials, can improve localization when a target is difficult to see directly. Their usefulness depends on a sufficiently stable relationship between marker and target. Placement is an invasive procedure and carries procedure-specific risks.
Relationship to other radiation techniques
IGRT and intensity-modulated radiation therapy (IMRT) have different functions. IMRT shapes the radiation dose by varying beam intensity, whereas IGRT checks anatomical alignment. The two are frequently used together.
Image guidance is particularly important in stereotactic body radiation therapy, where high doses are delivered in a small number of sessions and accurate localization is essential. It also supports motion-management techniques such as breath-hold, respiratory gating, and target tracking.
Adaptive radiation therapy goes beyond correcting position: it modifies the treatment plan in response to anatomical or other changes. Adaptation may occur between sessions or while the patient remains on the treatment table. IGRT can identify changes that prompt adaptation, but image guidance does not necessarily involve replanning.
Imaging also guides applicator placement and treatment planning in brachytherapy, although the term IGRT most often refers to external-beam workflows.
Clinical use and evidence
IGRT is used across many treatment sites, including prostate cancer, lung cancer, and head and neck cancers. The imaging method and frequency depend on target visibility, organ motion, treatment margins, dose per session, and available equipment.
Its immediate benefit is improved knowledge of treatment geometry. Whether a particular protocol improves tumor control, reduces side effects, or changes other clinical outcomes depends on the treatment setting. More accurate positioning alone does not establish a survival benefit, and evidence comparing imaging schedules and technologies is not uniform across cancer types.
Limitations and quality assurance
X-ray-based guidance adds imaging radiation exposure. This exposure is distinct from the therapeutic dose and varies with the modality, acquisition settings, anatomical region, and imaging frequency. MRI, ultrasound, and optical surface imaging do not add ionizing imaging exposure.
Other limitations include artifacts, poor target visibility, registration errors, additional treatment time, and the possibility that anatomy changes between imaging and irradiation. Positional correction cannot necessarily compensate for organ deformation or changes in the relationship between a tumor and surrounding tissues.
Reliable IGRT therefore depends on equipment commissioning, regular calibration, verification of alignment between imaging and treatment coordinates, staff training, and defined review procedures. These measures are part of patient safety: an inaccurate image match or incorrect correction can systematically misdirect otherwise precise radiation delivery.