The equivalence point is the stage of a titration at which the amount of added reagent, called the titrant, is chemically equivalent to the amount of the substance being determined, called the analyte. The relationship is defined by the stoichiometry of the specified reaction, not necessarily by equal volumes or equal numbers of moles. A central concept in analytical chemistry, the equivalence point provides the basis for calculating an unknown amount or concentration from the quantity of titrant required.
Stoichiometric basis
For a titration reaction represented as
where A is the analyte, B is the titrant, and and are their stoichiometric coefficients, equivalence occurs when
Here, is the initial amount of substance of analyte and is the amount of titrant delivered. These are analytical amounts; they do not imply that unreacted A and B remain present in those quantities at equilibrium.
When both substances are described by their molar concentrations, the relationship becomes
where is the titrant volume at equivalence. Thus,
For example, the complete neutralization of sulfuric acid by sodium hydroxide is represented by
At equivalence for complete neutralization, two moles of sodium hydroxide have been added per mole of sulfuric acid. Equivalence therefore does not generally mean equal molar amounts of the two reagents.
The definition presupposes a specified reaction. Competing reactions, impurities, or different possible reaction products can alter the amount of titrant consumed.
Equivalence point and endpoint
The equivalence point is a stoichiometric condition. The endpoint is the experimentally observed signal used to identify when a titration should stop or to estimate the corresponding titrant volume. The two are related but are not identical.
An endpoint may be indicated by:
- A persistent color change of a chemical indicator.
- A change in the potential measured using an electrode.
- A change in electrical conductivity.
- A change in optical absorbance.
- A change in temperature or another measurable property.
The difference between the endpoint volume and the equivalence-point volume is commonly expressed as a titration error:
For a direct titration with concentration calculated from the delivered volume, a positive produces an overestimate if no correction is applied. Endpoint selection aims to make this difference small and reproducible.
The equivalence point is not necessarily directly visible. It is inferred from a suitable chemical signal, a measured titration curve, or a model of the system.
Acid–base titrations
In an acid–base titration, equivalence occurs when the added acid or base matches the proton-transfer capacity of the analyte for the specified neutralization step. Its pH depends on the species present, their acid–base properties, concentration, solvent, and temperature.
Strong acid and strong base
For a strong acid titrated with a strong base in dilute aqueous solution, the principal reaction is
If the resulting salt does not appreciably affect acid–base equilibria, the solution is approximately neutral at equivalence. At , this corresponds to pH near 7 under ordinary dilute-solution conditions.
Neutrality does not always mean pH 7. It corresponds to equal hydrogen-ion and hydroxide-ion activities, so the neutral pH is
where is the ion product of water and . Its value changes with temperature.
Weak acids and weak bases
When a weak acid is titrated with a strong base, its conjugate base is present at equivalence and reacts with water:
This hydrolysis generally makes the equivalence solution basic. Conversely, titrating a weak base with a strong acid generally produces an acidic equivalence solution because the conjugate acid reacts with water.
Consequently, equivalence is not synonymous with neutral pH. The pH must be determined from the relevant chemical equilibria.
Half-equivalence and multiple equivalence points
For a simple monoprotic weak acid titrated with a strong base, the half-equivalence point occurs when half the stoichiometric titrant amount has been added. Under the usual approximations,
This is distinct from the equivalence point.
Polyprotic acids can have several equivalence points, corresponding to successive neutralization steps. Whether these appear as separate, clearly identifiable features depends on the separation between dissociation constants, concentration, and measurement method. A stoichiometric equivalence point can exist even when no distinct experimental endpoint is resolved.
Other types of titration
The concept is not restricted to neutralization.
Redox titrations. In a oxidation–reduction reaction, equivalence is determined by the balanced transfer of electrons. For example, permanganate reacting with iron(II) in acidic solution follows
One mole of permanganate is therefore equivalent to five moles of iron(II) under these conditions.
Precipitation titrations. Equivalence is reached when the titrant and analyte have been combined in the ratio required to form the precipitate. In the reaction of silver ions with chloride ions,
the ratio is 1:1. The finite solubility of silver chloride means that small dissolved concentrations remain at equivalence.
Complexometric titrations. Equivalence corresponds to the stoichiometric formation of a metal–ligand complex. Many titrations with EDTA use a 1:1 metal-to-EDTA ratio, although the completeness of complex formation and endpoint behavior depend strongly on pH and competing ligands.
Locating the equivalence point
A titration curve plots a measured property against added titrant volume. Around equivalence, a small addition may cause a large change in that property because the analyte’s capacity to consume further titrant has largely been exhausted.
For many potentiometric titrations, the equivalence volume is estimated from an inflection region, a maximum in the first derivative, or a zero crossing in the second derivative. These features do not coincide exactly with stoichiometric equivalence in every chemical system. Dilution, asymmetric equilibria, and the measured response can shift their positions.
Other methods use intersections of approximately linear portions of a conductivity curve or fit the data to an equilibrium model. Indicator-based methods instead depend on choosing a transition range within the steep part of the relevant titration curve.
Practical limitations and uncertainty
Reliable use of an equivalence point requires a known reaction stoichiometry, sufficiently complete reaction, adequate selectivity, and a reaction rate compatible with the measurement. Important limitations include:
- Incomplete reaction: an unfavorable equilibrium constant can produce a gradual rather than sharp transition.
- Slow reaction or mixing: delayed response can cause excess titrant to be added before the signal stabilizes.
- Competing reactions: other sample components may consume titrant.
- Poorly resolved steps: mixtures and polyfunctional substances may produce overlapping transitions.
- Instrumental effects: electrode response, delivery resolution, and calibration affect the estimated volume.
- Reagent changes: instability or contamination can alter the titrant concentration.
The measurement uncertainty of a titrimetric result therefore includes more than uncertainty in locating the endpoint. It can also include titrant standardization, sample preparation, delivered volumes, blank corrections, and the adequacy of the assumed chemical model.