{
  "id": "viscometer",
  "title": "Viscometer",
  "lang": "en",
  "source_lang": "en",
  "summary": "A viscometer measures a fluid’s resistance to flow or deformation, using methods such as timed capillary flow, rotational drag, falling objects, or vibrating sensors.",
  "category": "science",
  "subcategory": "physics",
  "type": "concept",
  "aliases": [
    "Viscosimeter"
  ],
  "generated_at": "2026-11-04",
  "model": "pro/gpt-6.1-sol",
  "url": "https://aiwiki.page/en/viscometer/",
  "translations": {
    "zh-Hans": {
      "url": "https://aiwiki.page/zh/viscometer/",
      "json": "https://aiwiki.page/zh/viscometer.json"
    }
  },
  "sections": [
    {
      "heading": "",
      "text": "A viscometer is an instrument used to measure viscosity, a fluid’s resistance to flow or deformation. It determines viscosity from a measurable response—such as flow time, pressure drop, rotational resistance, or damping of an oscillating sensor—under specified conditions. Viscometers are used in laboratory analysis, manufacturing, and process monitoring. Because viscosity depends on temperature and may also depend on the rate and history of deformation, a measurement is meaningful only when its operating conditions and method are identified."
    },
    {
      "heading": "What a viscometer measures",
      "text": "Dynamic and kinematic viscosity\n\nFor a Newtonian fluid in simple shear, dynamic viscosity \\(\\eta\\) relates shear stress \\(\\tau\\) to shear rate \\(\\dot{\\gamma}\\):\n\n\\[\\tau=\\eta\\dot{\\gamma}.\\]\n\nShear stress is tangential force per unit area; shear rate describes how rapidly adjacent fluid layers move relative to one another. A Newtonian fluid has a viscosity that is independent of shear rate at fixed temperature and pressure.\n\nIn the International System of Units, dynamic viscosity is expressed in pascal-seconds, \\(\\mathrm{Pa\\,s}\\). The millipascal-second, \\(\\mathrm{mPa\\,s}\\), is commonly used for liquids; it is numerically equal to the older centipoise unit.\n\nKinematic viscosity, \\(\\nu\\), is dynamic viscosity divided by density, \\(\\rho\\):\n\n\\[\\nu=\\frac{\\eta}{\\rho}.\\]\n\nIts SI unit is the square metre per second, \\(\\mathrm{m^2\\,s^{-1}}\\). The square millimetre per second, \\(\\mathrm{mm^2\\,s^{-1}}\\), is numerically equal to the centistoke. Gravity-driven glass capillary viscometers commonly determine kinematic viscosity, while many rotational instruments determine dynamic or apparent viscosity.\n\nNon-Newtonian measurements\n\nFor a non-Newtonian fluid, the ratio \\(\\tau/\\dot{\\gamma}\\) can vary with shear rate, elapsed time, or previous deformation. A viscometer may therefore report an apparent viscosity rather than a single material constant. Paints, concentrated suspensions, polymer solutions, and many foods exhibit such behavior.\n\nA rheometer generally offers broader control and characterization of deformation, including measurements of flow curves, yielding, and elastic response. The boundary between viscometers and rheometers is not absolute: a rotational instrument capable of measurements over several speeds may perform some rheological tests, whereas a simple viscometer may provide only one method-dependent value."
    },
    {
      "heading": "Main types and operating principles",
      "text": "Capillary viscometers\n\nCapillary viscometers measure flow through a narrow tube. For steady, fully developed flow of an incompressible Newtonian fluid through a circular capillary, the Hagen–Poiseuille equation gives\n\n\\[Q=\\frac{\\pi r^4\\Delta p}{8\\eta L},\\]\n\nwhere \\(Q\\) is volumetric flow rate, \\(r\\) is the tube’s internal radius, \\(L\\) is its length, and \\(\\Delta p\\) is the pressure difference across it. The relationship assumes laminar flow, negligible entrance effects, and no slip at the wall.\n\nIn gravity-driven glass instruments, the time \\(t\\) required for a specified volume to pass between timing marks is measured. Over the instrument’s applicable range, kinematic viscosity is often calculated as\n\n\\[\\nu=Ct,\\]\n\nwhere \\(C\\) is a calibrated instrument constant. Some conditions require additional corrections, including corrections for kinetic-energy effects at short flow times.\n\nCommon designs include Ostwald, Ubbelohde, and Cannon–Fenske viscometers. The suspended-level arrangement of an Ubbelohde instrument reduces sensitivity to the amount of liquid charged within its specified operating limits.\n\nPressure-driven capillary instruments can also characterize highly viscous materials, including polymer melts. Their interpretation may require corrections for entrance pressure losses, non-Newtonian flow, wall slip, and viscous heating.\n\nRotational viscometers\n\nRotational viscometers measure the torque required to rotate a spindle or measuring surface in contact with a sample, or measure rotation produced by an applied torque. Important geometries include:\n\n- Concentric cylinders: Fluid occupies the gap between an inner cylinder and an outer cup.\n- Cone and plate: A shallow cone rotates above a flat plate, producing an approximately uniform shear rate under ideal conditions.\n- Parallel plates: Fluid lies between two plates; shear rate varies with radial position.\n- Immersed spindles: A spindle rotates in a container, providing a practical measurement for routine quality control.\n\nWith a well-defined geometry and appropriate corrections, torque and angular speed can be converted to shear stress and shear rate. With some immersed-spindle arrangements, the flow field is less precisely defined, so the result depends strongly on spindle type, container dimensions, immersion depth, and rotational speed.\n\nRotational instruments are particularly useful for non-Newtonian fluids because measurements at different speeds reveal shear-dependent behavior. However, equal rotational speeds in different geometries do not necessarily produce equal shear rates.\n\nFalling-ball and rolling-ball viscometers\n\nFalling-ball viscometers infer viscosity from the speed of a sphere moving through a fluid. For an isolated sphere at terminal velocity in creeping flow, Stokes’ law gives\n\n\\[\\eta=\\frac{2r^2(\\rho_s-\\rho_f)g}{9v},\\]\n\nwhere \\(r\\) is sphere radius, \\(\\rho_s\\) and \\(\\rho_f\\) are sphere and fluid densities, \\(g\\) is gravitational acceleration, and \\(v\\) is terminal speed.\n\nThe equation assumes negligible inertial effects and a fluid extending far from the sphere. Real instruments require allowance for tube walls, end effects, and other departures from the ideal model. The Reynolds number helps establish whether creeping-flow assumptions are appropriate.\n\nInclined falling-ball and rolling-ball instruments use calibrated relationships between travel time, density difference, and viscosity. Their working equations are not necessarily the uncorrected free-fall Stokes expression.\n\nVibrational and oscillating viscometers\n\nThese instruments detect how a fluid loads and damps a vibrating element, such as a tuning fork, rod, wire, or torsionally oscillating surface. Viscosity is inferred from changes in amplitude, phase, resonance frequency, or the power needed to sustain oscillation.\n\nThe response depends on sensor design and may involve both viscosity and density. Consequently, some devices require an independent density value or an additional measurement to separate the two.\n\nVibrational instruments are useful for continuous process monitoring because they can be compact and have few moving parts. For non-Newtonian or viscoelastic samples, their readings may not be directly comparable with steady-shear measurements.\n\nEfflux cups and empirical instruments\n\nEfflux cups measure the time required for liquid to drain through a specified orifice. They are widely used for coatings, inks, and related products. Their direct result is a flow time; conversion to viscosity requires a suitable empirical relationship and a specified operating range.\n\nFlow time can be influenced by surface tension, wetting, and the changing liquid head. Efflux-cup results are therefore method-dependent and are not automatically interchangeable with capillary or rotational measurements."
    },
    {
      "heading": "Temperature control, calibration, and uncertainty",
      "text": "Temperature control is central to viscosity measurement. Liquid viscosity generally decreases as temperature rises, whereas dilute-gas viscosity generally increases. Even modest temperature differences can materially affect comparisons between samples.\n\nCalibration commonly uses reference liquids with assigned viscosities at specified temperatures. A complete measurement also depends on timer performance, dimensional accuracy, torque or pressure sensing, sample density when required, and the adequacy of the flow model. Metrological traceability concerns the documented calibration chain connecting a result to an appropriate reference.\n\nImportant sources of measurement uncertainty include:\n\n- Temperature gradients or insufficient thermal equilibration.\n- Bubbles, contamination, evaporation, or sample composition changes.\n- Timing, torque, pressure, or density measurement errors.\n- Incorrect filling, immersion depth, gap setting, or instrument alignment.\n- Wall slip, particle settling, entrance effects, or departures from the assumed flow regime.\n- Structural changes in the sample during the measurement.\n\nRepeatable readings do not by themselves establish accuracy: a stable instrument can still produce a systematic error if calibration or model assumptions are unsuitable."
    },
    {
      "heading": "Applications and interpretation",
      "text": "Viscometers support quality control of lubricants, fuels, paints, adhesives, cosmetics, foods, and pharmaceutical formulations. In chemical and materials research, they help characterize solutions and monitor changes during processing. Dilute-solution capillary measurements can also support estimates of polymer molecular size through appropriate material-specific relationships.\n\nIn-line instruments monitor flowing material directly, while laboratory instruments allow more controlled testing of collected samples. The two can yield different results when temperature, shear conditions, or sample history differ.\n\nFor Newtonian liquids, results from different valid methods should agree within their uncertainties when measured under equivalent conditions. For non-Newtonian materials, a reported viscosity usually needs additional information: instrument geometry, speed or shear rate, temperature, measurement duration, and sample preparation history. Without those details, an apparent difference in viscosity may reflect the measurement procedure rather than a difference in composition."
    }
  ],
  "links": [
    "viscosity",
    "temperature",
    "pressure",
    "international-system-of-units",
    "density",
    "non-newtonian-fluid",
    "torque",
    "reynolds-number",
    "surface-tension",
    "calibration",
    "metrological-traceability",
    "measurement-uncertainty"
  ],
  "sources": [],
  "license": "https://creativecommons.org/licenses/by/4.0/"
}