  "sections": [
    {
      "heading": "",
      "text": "The 10–20 system is an internationally standardized method for locating and naming scalp electrodes used in electroencephalography (EEG). Electrode positions are determined by distances measured over the head between identifiable anatomical landmarks. The name refers to intervals of 10% and 20% of these measured distances. By expressing positions as proportions rather than fixed lengths, the system accommodates differences in head size and provides a common framework for recording electrical activity associated with the brain."
    },
    {
      "heading": "Measurement principles",
      "text": "The system uses four principal anatomical landmarks:\n\n- Nasion: the depression at the root of the nose, near the junction of the frontal and nasal bones.\n- Inion: the prominent point of the external occipital protuberance at the back of the skull.\n- Left and right preauricular points: landmarks immediately anterior to the ears.\n\nMeasurements follow the curved surface of the scalp rather than straight lines through space. The principal measurements include a midline arc from nasion to inion and a transverse arc between the preauricular points. Additional circumferential and longitudinal measurements establish the remaining positions.\n\nAlong the nasion–inion midline, the sequence is:\n\nNasion → Fpz → Fz → Cz → Pz → Oz → inion\n\nThe first and last intervals are each 10% of the total arc length; the intervening intervals are each 20%. Fpz and Oz serve as measurement landmarks in the traditional layout but are not part of its usual set of 19 scalp recording sites.\n\nCz, the central midline position, is established using both the front-to-back and left-to-right measurements. The other positions are defined through related proportional measurements, producing an approximately symmetrical arrangement.\n\nThe percentages describe distances along specified scalp arcs. They do not mean that every pair of adjacent electrodes is separated by the same physical distance."
    },
    {
      "heading": "Electrode names",
      "text": "Each electrode label combines letters indicating an approximate scalp region with a number or midline designation.\n\nLabel | Meaning\nFp | Frontopolar\nF | Frontal\nC | Central\nT | Temporal\nP | Parietal\nO | Occipital\nz | Midline, from “zero”\n\nOdd numbers indicate the subject’s left side, and even numbers indicate the subject’s right side. Thus, F3 is a left frontal position, F4 is its right-sided counterpart, and Fz lies on the frontal midline. Left and right always refer to the person being examined, not the observer.\n\nThese labels describe scalp locations rather than precise subdivisions of the cerebral cortex. In particular, “central” is a topographic designation, not the name of a cerebral lobe. The correspondence between an electrode and underlying cortical structures varies with individual anatomy.\n\nTraditional electrode set\n\nThe conventional 10–20 array contains 19 scalp recording positions:\n\nRegion | Positions\nFrontopolar | Fp1, Fp2\nFrontal | F7, F3, Fz, F4, F8\nCentral and midtemporal | T3, C3, Cz, C4, T4\nParietal and posterior temporal | T5, P3, Pz, P4, T6\nOccipital | O1, O2\n\nTwo additional electrodes, commonly designated A1 and A2, are often placed at the left and right ears. Some recording arrangements instead use mastoid electrodes, commonly designated M1 and M2. Ear and mastoid sites are anatomically distinct and should not be treated as interchangeable descriptions.\n\nCounts of “10–20 electrodes” can therefore differ depending on whether ear, reference, ground, or supplementary electrodes are included.\n\nUpdated temporal labels\n\nIn expanded nomenclature, four traditional temporal labels have corresponding newer names:\n\nTraditional label | Updated label\nT3 | T7\nT4 | T8\nT5 | P7\nT6 | P8\n\nThe updated names make the original sites consistent with a larger coordinate scheme; they do not, by themselves, indicate different electrode locations. Both naming conventions remain present in clinical records, equipment, and publications."
    },
    {
      "heading": "Electrode placement and EEG montages",
      "text": "The 10–20 system specifies where electrodes are located. An EEG montage specifies which electrode signals are compared and how the resulting channels are displayed. These are separate aspects of recording.\n\nAn EEG channel ordinarily represents a voltage difference:\n\n\\[V_{\\text{channel}}(t)=V_A(t)-V_B(t).\\]\n\nIn a bipolar montage, a channel compares two recording electrodes, such as Fp1 and F3. In a referential montage, multiple electrodes are compared with a common reference. Digital recordings can often be displayed using different montages without moving the electrodes, provided the necessary signals were acquired.\n\nNo bodily recording site is completely electrically inactive. Consequently, reference selection affects the appearance of the EEG. Reference and ground electrodes also have different technical functions: the ground is not simply another name for the signal reference."
    },
    {
      "heading": "Historical development",
      "text": "The system emerged from efforts to standardize EEG recording during the mid-twentieth century. Earlier laboratories used differing placement methods and terminology, making recordings difficult to compare.\n\nHerbert H. Jasper’s 1958 report on the international ten–twenty electrode system became a foundational statement of the standard. Its combination of reproducible anatomical landmarks, proportional measurements, and consistent naming supported adoption across clinical laboratories and research institutions.\n\nThe basic arrangement subsequently provided the framework for denser electrode arrays rather than being entirely replaced by them."
    },
    {
      "heading": "Extensions and applications",
      "text": "The 10–10 system adds intermediate positions, using finer proportional spacing and an expanded naming scheme. Still denser arrangements include the 10–5 system. Such extensions increase spatial sampling while retaining relationships to the original landmarks and electrode positions.\n\nThe 10–20 framework is widely used in clinical EEG, particularly in the assessment of epilepsy, and in research involving sleep, sensory processing, and other aspects of brain function. Standardized positions support comparisons between examinations, communication between laboratories, and consistent descriptions of the scalp distribution of electrical activity.\n\nAn electrode location alone does not establish a diagnosis. Interpretation depends on the recorded patterns, their distribution and timing, recording conditions, and clinical context."
    },
    {
      "heading": "Limitations",
      "text": "The system standardizes scalp sampling, not the exact location of electrical generators within the brain. Electrical activity spreads through brain tissue, cerebrospinal fluid, skull, and scalp before reaching an electrode. A signal recorded at one site can therefore reflect contributions from multiple cortical regions.\n\nThe traditional 19-electrode arrangement provides relatively sparse spatial coverage, especially over inferior temporal and basal regions. Additional electrodes or denser arrays may reveal activity that is poorly represented in the basic layout.\n\nHead shape, landmark identification, placement accuracy, and cap fit affect the actual electrode coordinates. Standard labels therefore do not guarantee identical relationships to underlying brain structures across individuals. Research requiring more precise anatomical correspondence may use digitized electrode coordinates and individual magnetic resonance imaging.\n\nFinally, standardized placement does not eliminate recording artifacts. Eye movements, muscle activity, electrode movement, and electrical interference can affect recordings independently of whether electrodes occupy the correct 10–20 positions."
    }
  ],