  "id": "hounsfield-scale",
  "title": "Hounsfield Scale",
  "lang": "en",
  "source_lang": "en",
  "summary": "The Hounsfield scale expresses X-ray attenuation in computed tomography relative to water, supporting tissue characterization, image display, and quantitative analysis.",
  "category": "health",
  "subcategory": "medicine",
  "type": "concept",
  "aliases": [
    "CT attenuation scale",
    "CT number scale"
  ],
  "generated_at": "2026-11-05",
  "model": "pro/gpt-6.1-sol",
  "url": "https://aiwiki.page/en/hounsfield-scale/",
  "translations": {
    "zh-Hans": {
      "url": "https://aiwiki.page/zh/hounsfield-scale/",
      "json": "https://aiwiki.page/zh/hounsfield-scale.json"
    }
  },
  "sections": [
    {
      "heading": "",
      "text": "The Hounsfield scale is a standardized scale used in X-ray computed tomography (CT) to express how strongly a material attenuates X-rays relative to water. Its values are called CT numbers and are commonly reported in Hounsfield units (HU). Water is assigned 0 HU, and air is approximately −1,000 HU. Materials that attenuate X-rays more strongly than water generally have positive values; those that attenuate them less strongly have negative values. The scale is named after Godfrey Hounsfield, a pioneer of clinical CT."
    },
    {
      "heading": "Physical basis and definition",
      "text": "CT reconstructs a cross-sectional map of linear attenuation coefficients from X-ray measurements taken at multiple angles. The linear attenuation coefficient, conventionally written as \\(\\mu\\), describes the reduction in beam intensity per unit distance through a material.\n\nA commonly used definition of the CT number is\n\n\\[\\mathrm{HU}=1000\\,\\frac{\\mu_{\\text{material}}-\\mu_{\\text{water}}}{\\mu_{\\text{water}}}.\\]\n\nHere, \\(\\mu_{\\text{material}}\\) and \\(\\mu_{\\text{water}}\\) refer to attenuation coefficients under corresponding imaging conditions. This gives water a value of exactly 0 HU. Because air attenuates X-rays only very weakly compared with water, its value is approximately −1,000 HU.\n\nAlthough called a “unit,” the HU expresses a scaled, dimensionless ratio rather than an independent physical dimension. A difference of 1 HU corresponds to approximately 0.1% of the attenuation coefficient of water.\n\nCT numbers are not direct measurements of mass density. X-ray attenuation also depends on elemental composition and photon energy. Two materials with similar physical densities may therefore have different HU values, while different materials can produce overlapping values."
    },
    {
      "heading": "Representative values",
      "text": "Approximate values in conventional CT illustrate the scale:\n\nMaterial or tissue | Approximate CT number\nAir | −1,000 HU\nAerated lung | Commonly −900 to −500 HU\nAdipose tissue | Commonly −190 to −30 HU\nWater | 0 HU\nMany unenhanced soft tissues | Approximately +20 to +80 HU\nDense cortical [[bone | bone]]\n\nThese ranges are illustrative, not fixed biological constants or diagnostic thresholds. Lung attenuation varies with inflation and the proportions of air, blood, and tissue. Bone values depend on mineral content, structure, and scanning conditions.\n\nIodinated contrast material can substantially increase the attenuation of blood vessels and organs. Metals may produce values of several thousand HU and can also cause artifacts that distort measurements nearby. The upper limit of the scale depends on the image format and reconstruction implementation; +1,000 HU is not a universal maximum."
    },
    {
      "heading": "Image display and measurement",
      "text": "Each voxel, or three-dimensional image element, has a reconstructed CT number. On a two-dimensional image, that value is represented by a pixel.\n\nA CT image usually contains a wider range of HU values than can be usefully distinguished in a single grayscale display. Windowing selects which values are mapped onto the visible shades:\n\n- Window level specifies the center of the displayed range.\n- Window width specifies the span of CT numbers distributed across the grayscale.\n\nA narrow window emphasizes small attenuation differences, whereas a wide window accommodates a larger range. Different window settings are therefore used to display lung, soft tissue, and bone. Changing the window changes the appearance of the image, not its underlying CT numbers.\n\nMeasurements are commonly obtained by drawing a region of interest over a structure. Software can report its mean CT number and the variation within it. Measurements near boundaries may combine several materials and may not represent any one tissue accurately."
    },
    {
      "heading": "Applications",
      "text": "In radiology, attenuation measurements complement visual assessment of anatomy and image patterns. They help identify fat, fluid, mineralization, gas, and contrast enhancement, although overlapping values mean that an HU measurement alone rarely establishes a diagnosis.\n\nComparison of unenhanced and contrast-enhanced scans can quantify changes in attenuation. Such comparisons depend on acquisition timing, contrast administration, and consistent measurement technique.\n\nQuantitative applications include assessment of lung aeration, body composition, and urinary stone attenuation. Estimates of bone mineral density may also use CT measurements, but require suitable calibration or validated methods rather than treating HU as an interchangeable measure of mineral density.\n\nIn radiation therapy planning, CT numbers are converted into quantities such as relative electron density or material properties used in dose calculations. These conversions require appropriate scanner and protocol calibration."
    },
    {
      "heading": "Limitations and sources of variation",
      "text": "The Hounsfield scale provides a common reference, but it does not make all CT measurements directly interchangeable.\n\nX-ray energy. Attenuation depends on photon energy. Changing tube voltage or the X-ray spectrum can change HU values, especially for materials containing elements such as iodine or calcium. Conventional CT uses a spectrum of energies, so measured values are not simply material constants at one photon energy.\n\nReconstruction and noise. Reconstruction algorithms, filtering, slice thickness, radiation exposure, and patient size influence noise and may affect measurements. Small differences between scans can reflect technical variation rather than a change in tissue.\n\nPartial-volume effects. A voxel containing both tissue and air, or both soft tissue and bone, yields a combined value. This particularly affects small structures and measurements near edges.\n\nArtifacts. Beam hardening, scattered radiation, motion, and metal-related artifacts can alter reconstructed CT numbers. A low or high value in an artifact-affected region may not reflect the actual material.\n\nCalibration. Routine calibration and quality-control testing use reference materials, including water, to check CT-number accuracy and uniformity. Even with calibration, measurements require attention to measurement uncertainty and acquisition conditions.\n\nSpectral and dual-energy CT can produce energy-specific images and material maps. Their outputs require distinction from conventional HU images: an iodine concentration map, for example, represents a different quantity. Likewise, grayscale values from magnetic resonance imaging are not ordinarily Hounsfield units."
    }
  ],
  "links": [
    "x-ray-computed-tomography",
    "water",
    "density",
    "bone",
    "contrast-agent",
    "voxel",
    "radiology",
    "diagnosis",
    "kidney-stones",
    "radiation-therapy",
    "calibration",
    "measurement-uncertainty",
    "magnetic-resonance-imaging"
  ],
  "sources": [],
  "license": "https://creativecommons.org/licenses/by/4.0/"
}